Electron Interaction Volume

Schematic drawing of electron interaction volume.

As electrons enter the sample they are scattered by the atoms in the sample into what is called the interaction volume. Characteristic x-rays produced from the sample are generated within this volume. The size of the interaction volume is determined by the density of the sample and by the energy of the electron beam. In a typical geological sample with a beam energy of 15 thousand volts the interaction volume is approximately 2 micrometers in diameter. Even if the diameter of the electron beam entering the sample is much smaller than a micrometer, the electrons will be scattered and the lateral resolution of the analysis will be limited to the size of the interaction volume.


Return to How Does the Electron Microprobe Work? | Return to Electron Microprobe Home | Return to Denver Microbeam Laboratory Home


Commercial equipment is identified at this web site for clarity and to properly define laboratory capabilities. Such identification does not imply recommendation or endorsement by the USGS, DOI, or the U.S. Government, nor does it imply that the equipment identified is necessarily the best available for the purpose.


Content information: Greg Meeker

To skip the program navigation area click here. Mineral Resources Eastern /  Central /  Western /  Alaska /  Minerals Information
Crustal Imaging & Characterization /  Spatial Data

Accessibility FOIA Privacy Policies and Notices

Take Pride in America logo USA.gov logo U.S. Department of the Interior | U.S. Geological Survey
URL: http://usgsprobe.cr.usgs.gov/eprobe8.html
Page Contact Information: Webmaster
Page Last Modified: Tuesday, 16-Oct-2007 13:05:47 MDT